167 posters, 28 videos, 9 topics, 12 sessions, 337 authors, 160 institutions
ePostersLive by SciGen Technologies S.A. All rights reserved.
ISMICS 2023 Annual Meeting
31 May - 3 June, 2023 | Boston, MA

PC32
Sternal Closure Using Rigid Plate Fixation: Design Improvements For Enhanced Patient Recovery
Poster Presenter
Authors
Affiliations
St. Luke's Hospital, The Mid America Heart Institute, St. Luke's Hospital, The Mid America Heart Institute, Kansas City, MO, Franciscan Health, Indianapolis, IN, The University of Chicago, Chicago, IL, New York Presbyterian Hospital, Columbia University Medical Center, New York, NY
Part of Topic
Education, Imaging and Facilitating Technologies
Objectives
In a multicenter randomized trial, sternotomy closure using rigid plate fixation (RPF) showed1,2
This bench top study evaluated the biomechanical properties and innovations of a next generation RPF system.
Methods
Biomechanical bench testing in a simulated sternotomy model compared the threshold for failure of an RPF that incorporates an integrated amorphous cable (SternaLock XP, Zimmer Biomet, Jacksonville, FL) to the legacy plating system (SternaLock Blu, Zimmer Biomet, Jacksonville, FL). The amorphous band increases surface area and resist pull through (Figure 1).
Results
SternaLock XP with an integrated amorphous cable cerclage performed significantly better and was 50% stronger compared to the legacy RPF system alone (Fig 3). SternaLock EZ compared to the legacy RPF system was 15% stronger and should be easier to use with faster implants and better screw engagement through its enhanced implant instruments and features (Fig 4).
Conclusions
Design enhancements over a current legacy RPF system which incorporate an integrated amorphous cable with the plate