<p>Evaluation of Machinability of Feldspathic Blocks Using ISO 18675 Merlon Test</p> <p> </p>

Part of Topic

Original Research

Objective:

To evaluate the minimum required thickness for the success of milled ceramic restorations with a standardized method of preparation and milling procedure and to compare milling damage of Mark II specimens fabricated using two milling units: Primemill and MC X5

Materials and Methods: Fifty (50) VITA Mark II CAD/CAM blocks (size of I-14 and shade 3M1C) were used to fabricate merlon specimens using Sirona inLab MC X5 and CEREC PrimeMill with specifications according to ISO 18675. Five specimens of each thickness of 0.1, 0.2, 0.3, 0.4, and 0.5 mm were prepared with each milling machine. For VITA mark II, lot number was 95420 and a reference no. of EC43M1CI14 and of March 2023 manufacturing date

 

In Cerec Primemill, tools set were extra fine grinding diamond burs (white colored: sizes 1.0 CS, 0.6 CS, 1.4 CS, and 1.2 CS) with a normal speed were used. Milling parameters were inputted by the CAM software of CAD/CAM system. Wet milling was used

 

Some burs were changed before the third block

milling of group 0.4 mm thickness as the screen

showed lower percentage level of function and they

were as follows: bur 1.2 mm had remaining percentage

of 66% and bur 1.4 mm had 76% of its 100%. There was

another burs changing process before the milling of

the fifth block of 0.1 mm thickness and they were: 1.2 mm bur with 55% remaining

 and 1.4 mm bur with 68% remaining

 

The burs used in MC X5 were diamond grinders or diamond coated grinders. Diamond grinders have a white color-coded size sequence of 0.6, 1.2, 1.4, and 2.2 mm, whereas all others have a yellow color-coded sequence of 0.5, 1.0, and 2.5 mm

 After the milling preparation process in Cerec PrimeMill and inlab X5, the merlon walls and bottoms were measured for thickness

A merlon wall was considered failed if more than one third (1/3) of its top end was not complete seen by naked eye. The bottom was considered fractured if a hole can be visually observed

 

 

 

Measurements were taken using a Mitutoyo caliper (0.02 mm tolerance) for wall thickness and micrometer (0.001 mm tolerance) for bottom thickness.

 

 

 

Results:  The minimal machine thickness (with a 100% success rate) for the Merlon wall is 0.2 mm for MC-X5 and 0.4 mm for PrimeMill. The minimal machine thickness (with a 100% success rate) for the bottom is 0.5 mm for MC X5 and 0.2 mm for PrimeMill milling machines

 

 

 

Conclusion:

 

Wall Thickness Success Rate in MC X5 versus Primemill

 