Loading...

501 posters, 2 videos, 1 audios, 8 topics, 37 sessions, 395 authors, 307 institutions

ePostersLive by SciGen Technologies S.A. All rights reserved.

23rd Annual Pain Medicine Meeting 2024

November 21-23, 2024 | Caesars Palace, Las Vegas, Nevada

23rd Annual Pain Medicine Meeting 2024 banner

5738

UNRELIABLE USE OF A CONTINUOUS GLUCOSE MONITOR IN THE OUTPATIENT PROCEDURAL SETTING

Part of Topic

Safety/QA/QI Projects

Unreliable Use of a Continuous Glucose Monitor in the Outpatient Procedural Setting

Evan Reuter, MD Ochsner Pain Medicine Fellow

Whitney Braddy, MD Ochsner Pain Medicine Fellowship Associate Program Director

  • Glucocorticoid injections pose risks to diabetic patients due to potential blood glucose elevation.

  • As CGMs become widespread, it may be tempting to defer facility testing and rely on patients' home monitors.

  • Reliable inpatient use of CGMs has been reported.

  • Used in COVID units

  • CGMs used inpatient required frequent calibration by nursing

  • FDA Guidelines

  • Known variance between capillary glucose and serum glucose

  • Permits CGM vs serum laboratory measures a difference of 20%.

  • Facility glucometers are calibrated every morning.

  • Home CGMs are intended to be calibrated every 10 days.

 
  1. Patient presented for scheduled procedure

  2. Patient has point of care glucose check due to diabetic status using facility POC capillary glucose test

  3. Facility glucometer presents reading of 244 mg/dL

  4. We informed patient of need to postpone procedure

  5. Patient presents home CGM reading of 142 mg/dL. See figure 1.

  6. Repeat facility glucometer readings are congruent with hyperglycemia.

  7. Procedure is postponed.

 
We present an episode of an unreliable home CGM (continuous glucose monitor) at the time of an outpatient procedure. We share this case for all proceduralists to be aware that home CGMs cannot be trusted in procedure safety decisions.