EVALUATING VERIFICATION MODALITIES FOR EPIDURAL CATHETER PLACEMENT

Poster Presenter

Aaron Berg

Part of Topic

Scientific Abstracts > Emerging Technology

  • Evaluating Verification Modalities for Epidural Catheter Placement

  • Epidurogram is the current gold standard for confirmation of epidural catheter placement. 

    • Involves delivering both contrast dye and radiation to the patient 

  • Compuflo CathCheck is a non-invasive system utilizing pulse wave analysis to determine catheter location with minimal risk to the patient

  •  Determine if Compuflo CathCheck system can accurately confirm the correct placement of a thoracic epidural catheter, verified with epidurogram

  •  Other methods of correct epidural catheter placement were also evaluated, including meniscus test and physical exam

  • Single center crossover study at the University of Minnesota 

  • After IRB approval, informed consent was obtained on all patients prior to their epidural placement, per standard protocol 

  • Each patient underwent 4 different catheter detection modalities: meniscus test, sensory examination with ice, Compuflo CathCheck (Milestone Scientific,Livingston, NJ), and epidurogram 

  • All exams were performed or interpreted by an individual who was blinded to the other tests

  • The results of all testing was sent to a research assistant that compiled and entered all data.

  • All testing was validated with the epidurogram

  • Correct epidural catheter placement was confirmed via epidurogram in 16 subjects

  • Incorrect placement, outside of the epidural space, was confirmed in 3 subjects

  • CompuFlo CathCheck system had 100% agreement with the results of epidurogram for both correct and incorrect placement

  • Physical Exam was in agreement with epidurogram 56% of the time for correctly placed epidurals, 33% of the time for incorrect placement

  • Meniscus Test was in agreement with epidurogram 63% in correctly placed epidurals and 33% in incorrect placement

  • The CompuFlo CathCheck system appears to accurately determine correct and incorrect epidural catheter placement for thoracic epidurals when compared to epidurogram

  • Neither physical examination nor meniscus testing appear to accurately reflect correct epidural catheter placement